OREC™ 0500 – 0900 Series Ozone Test Chambers: Specifications
model: 0500 0900 0500A 0900A 0500C 0900C
test methods: 1 c · d · g · h
j · i · k · l
a · e · f · g
h · i · j · k · l
a · c · d · g· f
h · i · j · k · l
ozone generator: quartz uv corona discharge quartz uv &
corona discharge
ozone range: 0–250 pphm/vol

0–1000 pphm/vol 2
10–300 ppm/vol 0–250 pphm/vol
10–300 ppm/vol
0–1000 pphm/vol 2
air flow: cfm 0–6 0–15 10–20 10–20 10–20
0–61
10–20
0–15
system accuracy: measurement: 3% of reading      Control: 4% of set–point
temperature: up to 90° C (194º F)
oven chamber: cm
in.
61 x 46 x 51
24 x 18 x 20
76 x 76 x 76
30 x 30 x 30
61 x 46 x 51
24 x 18 x 20
76 x 76 x 76
30 x 30 x 30
61 x 46 x 51
24 x 18 x 20
76 x 76 x 76
30 x 30 x 30
oven chamber: m³
ft³
0.142
5.000
00.443
15.625
0.142
5.000
00.443
15.625
0.142
5.000
00.443
15.625
dimensions: cm
in.
122 x 64 x 84
48 x 25 x 33
137 x 91 x 114
54 x 36 x 45
122 x 64 x 84
48 x 25 x 33
137 x 91 x 114
54 x 36 x 45
122 x 64 x 84
48 x 25 x 33
137 x 91 x 114
54 x 36 x 45
net weight: kg
lb
180
410
261
580
273
600
355
780
282
620
364
800
Information1 test method references
a) ASTM D470 crosslinked insulations and jackets for wire and cable
b) ASTM D518 rubber deterioration–surface cracking
c) ASTM D1149 rubber deterioration–surface ozone cracking in a chamber
d) ASTM D1171 rubber deterioration–surface ozone cracking outdoors or chamber
e) ASTM D1352 ozone–resisting butyl rubber insulation for wire and cable
f) ASTM D1373 specification for medium-voltage rubber insulating tape: withdrawn in 1986
g) ASTM D3041 testing coated fabrics–ozone cracking in a chamber: replaced by D1149
h) ASTM D3395 rubber deterioration–dynamic ozone cracking in a chamber
i) ASTM D4575 rubber deterioration reference and alternative method(s) for determining ozone level in laboratory test chambers, (method a only)
j) ISO 1431–1: 1989 resistance to ozone cracking – part 1: static strain
k) ISO 1431–2: 1994 resistance to ozone cracking – part 2: dynamic strain test
l) ISO 1431–3: 2000 resistance to ozone cracking – part 3: reference and alternative methods for determining the ozone concentration in laboratory test chambers
Information 2 requires the optional additional uv lamp.

| quickquote | techinfo | adobe® download | versión española | back | home |

© 2006 Ozone Research Equipment Company • All Rights Reserved • Published: June 20, 2006
221 Beaver Street • Akron, Ohio 44304 • U.S.A.
Telephone: 800.742.8535 • Facsimile: 800.229.9329